7625 West New York Street
Indianapolis, Indiana 46214-4911 USA
Telephone: (317) 273-6960
Facsimile: (317) 273-6979
Order Desk Telephone: (800) 401-1723
Order Desk Facsimile: (800) 515-9254
E-mail: flotec@floteco2.com

Home Page  

Products Page

 

Vortran Automatic Resuscitator (VAR)

 

Critical Care       

                                                       VORTRAN Automatic Resuscitator™

                                              

 

 

 

 

     

 

              Model VOR-R12 Shown                   Model VOR-R11 Shown

Gas-powered automatic resuscitator for emergency ventilation 

available for immediate shipment.

The VORTRAN Automatic Resuscitator™ (VAR) is a unique single patient use, disposable resuscitator. It provides consistent, reliable, hands free ventilatory support via a mask or endotracheal tube using a continuous gas flow source (Model RT - a.k.a. the original RespirTech PRO). The Model RC adds a gas entrainment feature for the delivery of 50% FiO2 by entraining room air, significantly reducing the consumption of oxygen. The VARs are to be used by properly trained personnel for the delivery of short term, constant flow-pressure cycled ventilatory support in emergency and hospital environments. They are cost competitive and provide more consistent ventilation than manual resuscitators

Optional Delivery of FiO2 at 50% or 100%. At 50% Cylinder usage time is more than doubled. 

Consistent hands free ventilatory support  • Automatic pressure cycled ventilation

For breathing or non-breathing patients    Disposable - single patient use.

Part #

VOR-R00

 

 

7 ft. supply tube, 6" flex hose

cs / 10

Part #

VOR-R01

M

 

7 ft. supply tube, 6" flex hose

cs / 10

Part #

VOR-R02

M

 

20 ft. supply tube, 6" flex hose, 5' 22mm flex hose

cs / 10

Part #

VOR-R03

M

 

VOR-R01 4-pack

bx / 4

Part #

VOR-R10

 

 E

7 ft. supply tube, 6" flex hose

cs / 10

Part #

VOR-R11

M

E

7 ft. supply tube

cs / 10

Part #

VOR-R12

M

E

20 ft. supply tube, 6" flex hose, 5' 22mm flex hose

cs / 10

Part #

VOR-R13

M

 

7 ft. supply tube, 6” flex hose   (Peds 10 kg and higher)

cs / 10

E – includes FiO2 gas entrainment adapter                                     M – includes Manometer

Pediatric Model Available

VORTRAN Automatic Resuscitator™ (VAR). Unique single patient, multiple-use disposable emergency resuscitator.

How does VAR function?

The gas-powered VAR provides constant flow, pressure cycled automatic ventilatory support for both breathing and non-breathing patients. The primary working mechanism of the VAR is the [1] modulator with [a] peak inspiratory pressure (PIP) and [b] breathing rate adjustment dials, which is an exhalation valve that opens at one pressure (PIP) and closes at another lower pressure (PEEP). The remaining components of the VAR consist of the [2] pressure manometer, [3] inlet for supply gas flow, [4] patient connection port, [5] redundant pressure pop-off valve, and [6] one-way valve for entraining additional air (Figure 1).

The pulmonary modulator provides the actual ventilatory support. The primary working mechanism of the pulmonary modulator is the piston. The piston is spring loaded, designed like a pressure pop-off valve except the spring force is adjustable (the [a] PIP Dial).

Figure 1
VAR Component Description

 

Figure 2
VAR Dimensional Profile

Clinical Considerations

The VAR provides short term, pressure cycled, and constant flow ventilatory support for either breathing or non-breathing patients. This allows the patient to receive consistent, and reliable ventilatory support. The VAR is pressure cycled on inhalation and exhalation (PIP and PEEP) which minimizes the possibility of gas trapping. During inhalation, exhalation will not start until PIP is reached. During exhalation, inhalation will not begin until pressure drops to PEEP. For spontaneously breathing patient, the rate dial of the VAR is set so that the base line pressure is above the intrinsic PEEP allowing the patient to initiate inhalation by drawing the base line pressure down to the set PEEP. Because the VAR is a constant flow pressure cycled device, changes in patient compliance will result in changes in the respiratory rate (stiffer or smaller compliances produce faster rates). The advantage of this minimizes the danger of barotrauma. However, it should be emphasized that the VAR is to be used only by trained personnel who continuously monitor the patient. The VAR is not an ICU stand-alone ventilator with multiple monitoring features.

Setup and use of the VAR is simple (refer to Setup Instructions in Section 3). Set desired flow (Q), adjust PIP pressure dial to obtain desired inspiratory time (ti) to attain tidal volume (TV = Q X ti), see Tidal Volume Table below. The gas flow, patient's lung compliance and PIP settings control the inspiratory time and tidal volume. Then adjust rate dial to obtain desired breathing rate.

Table
Estimated Tidal Volume (mL) at Various Flow Rate (LPM)

The VAR runs on a continuous gas flow (inspiratory flow) of 15 to 40 L/min depending on patients' inspiratory flow demand. When connected to a 50 PSIG gas source, the VAR will automatically deliver 40 L/min (667 mL/second) per ASTM guideline[1]. Delivered tidal volume may be determined by multiplying the flow in mL/second and the inspiratory time in seconds, or using the estimated tidal volume table.

The rate dial controls exhalation time (t
e), and when dialed down enough will cause the VAR to stop cycling automatically (infinite exhalation time). Under these circumstances the VAR is delivering pressure supported ventilatory support and the patient must trigger the VAR to begin subsequent full inhalations. If the patient is apneic or pressure control ventilation is desired restart automatic cycling of the VAR by adjusting the rate dial counter clockwise until cycling begins again. Whenever the VAR stops cycling, the first step, in the absence of obvious clinical factors, is to check if it is in pressure support mode by rotating the rate dial counter clockwise (out). If rotating the rate dial counter clockwise substantially (3 or 4 turns) does not start automatic cycling, the patient's airway may be occluded or a very large leak exists.

PIP may be adjusted from 20 and 50 cm H
2O. PEEP is typically 1/10th of PIP. Inspiratory time and rate are adjustable over a wide range. Changes in the PIP setting or flow will also affect the respiratory rate. It is important to check all settings when making a change to any of these three variables (flow, PIP and rate). For example: reducing the PIP setting may cause the VAR to go into spontaneous breathing mode. Adjust the rate dial out (counter-clockwise) to restart automatic cycling.

The VAR is equipped with an air entrainment valve which allows the patient to entrain additional air and respond to the demands of the patient. Patient entrainment of outside air is normally audibly detectable and the percent oxygen delivered to the patient will be reduced. Specific concentrations of oxygen may be delivered to the patient with the use of an oxygen blender.

Although the design of the modulator is similar to that of a pop-off valve and is inherently safe, the VAR is also equipped with a redundant pop-off valve that relieves pressure at 60 cm H
2O. When the pop-off valve is activated, the pop-off valve piston will be seen to open slightly and excess pressure released.

The duration of an "E" cylinder when using a VAR will depend on the flow. An "E" cylinder contains 625 L of gas. At 40 L/min, 625 L will last up to 15 minutes; at 20 L/min, 625 L will last up to 30 minutes. 15 L/min orifice type flowmeters used on many "E" cylinders will not be able to deliver more than 15 L/min. When clinicians decide that 15 L/min is not sufficient flow, the VAR can be attached to a regulator that has a high flow port (50PSIG) to deliver 40 L/min.

Although peak pressures are listed on the side of the pressure dial, they are only approximate. Clinicians using the VAR are still required to use good clinical judgment and monitor the patient appropriately. A manometer may be connected between the modulator and the patient connector tee.

The VAR is pressure cycled on PEEP as well as PIP. In the pressure control mode there is no prolonged stage where the flow of exhalation gas stops for a significant duration of time (in the pressure support mode, exhalation time is determined by the patient). This occurs because the exhalation time is set with the rate dial by varying the exhalation resistance so that the patient just finishes exhalation with the beginning of the subsequent inhalation. The volume of gas with which the patient's lungs are inflated with when reaching PEEP is the same as with any other means of obtaining PEEP. As with all ventilatory support modes, short exhalation times on patients with high airway resistance may lead to gas trapping which is not detectable in the patient's external airways. Upon occlusion of the patient's airways, the VAR will stop cycling or may sometimes cycle rapidly.

The VAR will work with any mask that provides a good seal with the patient. All clinicians should receive adequate training with a VAR with mask prior to use. In the presence of a small leak, the VAR will still cycle between PIP and PEEP. Noticeable changes in the presence of a leak are increased inspiratory times and decreased expiratory times. The VAR works very well with an endotracheal tube.

Inhalation may be initiated by briefly removing the mask from the patient or briefly disconnecting the modulator from the patient adapter tee. In either event, inhalation begins because pressure has dropped down to PEEP and the VAR is pressure cycled.

Upon contamination of the VAR with vomitus, it may be cleared by disconnecting the modulator from the patient connector tee (see enclosed instructions) and tapping out vomitus on a hard surface. Additionally, if needed, the rate dial may also be removed to facilitate removal of vomitus from modulator. This operation should take less then 20 seconds, and in a lab setting has consistently been shown to take approximately 11 seconds. Alternatively, upon contamination with vomitus, the clinician may choose to discard the device and use a new one.

Inhalation and exhalation are audibly detectable and easily recognizable during operation of the VAR.

The VAR may be controlled remotely by connecting any length of 22-mm corrugated tubing between the patient connector tee and the modulator. The attached tubing will not increase the dead space, the modulator is an exhalation control valve, inspiratory gas is delivered through the patient connector tee.

The primary advantage of the VAR, as compared to manual resuscitators is the ability to deliver consistent, reliable, hands free resuscitation. Manual resuscitators may have adverse effects on patients as a result of inconsistent ventilation (see Clinical Reference in Section 5).

Figure 3

 

1.  PIP-Set by PIP Dial, control Inspiratory TIME (Itime)
2.  PEEP _ Approximately 1/10th of PIP setting
3.  Inspiratory Flow Rate (Q) _ Maximum 40 L/min (= 667 mL/sec)
4.  Inspiratory Time (Itime) _ Time required to reach PIP
5.  Exhalation Time (Etime) _ Time required to drop from PIP to PEEP
6.  Tidal Volume = Q X Itime
7.  Respiratory Rate (RR) = 60 / (Itime + Etime )
8.  RATE DIAL _ Set exhalation resistance and change RR


[1] Standard Specification for Minimum Performance and Safety requirements for Resuscitators Intended for Use with Humans, ASTM Designation: F 920 _ 93.