![]() |
||
| 7625 West New York Street | ||
| Indianapolis, Indiana 46214-4911 USA | ||
| Telephone: (317) 273-6960 | ||
| Facsimile: (317) 273-6979 | ||
| Order Desk Telephone: (800) 401-1723 | ||
| Order Desk Facsimile: (800) 515-9254 | ||
| E-mail: flotec@floteco2.com | ||
TECHNICAL BULLETIN
MRI (MAGNETIC) SAFETY & COMPATIBILITY OF FLOTEC ALUMINUM REGULATORS
MRI Safe and MRI Compatible.
The following is the test report provided by:
Emanuel Kanal, MD, FACR Director, MR Services UPMC Presbyterian 200 Lothrop Street Department of Radiology, Room D-132 Pittsburgh, PA 15213-2582Methods:
The Flotec oxygen regulator submitted for testing was tested for deflection. The oxygen regulator was also tested for torque and artifact generation potential. All testing was performed in a GE Signa® LX 3.0T MR System. Specifications: multi-coil super conducting, actively shielded magnet housed in a 94cm bore; gradient amplitude - 40mT/m and slew rate - 150T/m/s; 8.2.5 M4 software. Using gauss line plots and visual inspection, the location of where the spatial magnetic field gradients were the greatest was determined. This location was marked by color-coded tape.
Using established guidelines*, an apparatus was constructed to hold the implants for testing. This apparatus consisted of a plastic protractor with a string to hold the implants and a plastic level glued to the top of the protractor. The protractor assembly was attached to a slide mechanism on a plastic rod that was attached to a flat piece of Plexiglas. The protractor was placed far enough away from the rod to allow for free movement of the implants. Using the scanner alignment lights, isocenter in the coronal plane was marked on the rod.
The oxygen regulator was tested with a hand held magnet prior to measurements. No gross ferromagnetic attraction was noted with the oxygen regulator when exposed to the relatively weak hand magnet, and we proceeded with the formal phase of the testing.
The oxygen regulator was weighed on a top loading balance, XL-5K, Denver Instrument Company. The string used to hold the oxygen regulator was weighed using a top loading balance, XL-300, Denver Instrument Company.
All sources of forced air movement were shut off. This included the patient fan inside the bore of the magnet, as well as the gradient cooling fan. The apparatus was aligned in the direction of the main magnetic field and centered in the sagittal plane using the scanner alignment lights. The device was attached to the string and placed at isocenter in the vertical axis as determined by the MR scanner alignment lights. The protractor was leveled horizontally. A landmark was then established. The apparatus was advanced to the location of the maximum gradient change marked previously. The table location, as indicated by the LED display, was documented. After steadying the implant, with the string at 90 degrees, the device was released to move freely and the deflection angle was recorded. Three separate measurements were taken and recorded.
Torque was measured on this single device by positioning it perpendicular to the orientation of the MR scanner's magnetic field/magnetic lines of force while on a sheet of plate glass at magnet isocenter and grossly observing for any rotational realignment forces attempting to reestablish a parallel orientation.
Flow measurements were examined on the Flotec oxygen regulator, S/N 421763 model in two separate locations: 1) Location #1 just inside the magnet bore opening, and 2) Location #2 at the entrance to the magnet room (approximately 33 feet away from the magnet bore opening). Measurements were taken with an attached Sherwood MR compatible post valve (tested by our facility at 3 Tesla) in the "On", or fully opened, position. Flows were set at the oxygen flow rate control unit at the side wall of the MR scanner and measurements called off by a second individual blinded to the flow settings at the wall unit control location. Flow measurements were made using a standard flow meter positioned >10 feet from the magnet bore entry location. Of that which was set on the single oxygen regulator so tested (Flotec oxygen regulator, S/N 421763) setting
To test for possible artifacts within MR images obtained while this oxygen regulator was in use, two separate scans were performed to determine if the presence of the oxygen regulator in the magnetic field created any degradation in the MR images. The GEMS nickel doped spherical phantom, in the loader, was scanned, using gradient echo sequences with an echo time of 20 milliseconds, flip angle of 45 degrees, repetition time of 200 milliseconds, 24cm x 24cm field of view, 5mm slice thickness with a 1mm gap. This sequence was performed twice; once with the post valve/oxygen regulator located at the bore entry location and then again with these devices removed from the bore/bore entry position.
*Standard Test Method for Measurement of Magnetically Induced Displacement Force on Passive Implants in the Magnetic Resonance Environment, 6. Apparatus
Results:
The Flotec Oxygen Regulator S/N 421763 demonstrated 21 degrees of deflection on the deflection angle test; this is below the 45-degree threshold necessary for claims of MR safety/compatibility. Torque test was grossly negative for this model/device. No significant artifacts were observed in the image with this oxygen regulator/attached post valve positioned at the magnet bore entry location. Signal to noise measurements of the phantom with the above -noted post valve still attached to the Flotec Oxygen Regulator, S/N 421763 at the bore entry location, were comparable (150.6 for the baseline and 152.6 with the devices at the bore entry location) and well within one standard deviation of noise measurements (noise measurements were 23.18 for the baseline and 22.8 with the post valve/oxygen regulator at the bore entry location).) Flow rates measured at increments of <=1 l/m throughout the range of 0 to 6 l/m at both locations #1 and #2 measured consistently within roughly 0.5 l/m of that set on the wall control unit. Thus, no alteration of flow rate/function was identified or observed for this oxygen regulator S/N 421763 that appeared in any way dependent upon or modified by the presence or absence of the static magnetic field and static spatial magnetic field gradients of the 3T MR scanner.
Conclusions:
It is my opinion that the present submitted Flotec Oxygen Regulator S/N 421763 model tested does meet the criteria for both MR safety as well as MR compatibility at 3 Tesla when used up to and including at the magnet bore entry position of this system on which it has been tested.
Please note that grossly detectable Lenz's Law related forces when torqued at bore entry and even greater such detectable forces at magnet isocenter are expected and predictable for metallic objects of the mass/geometry, and should not be misconstrued as affecting present definitions of product labeling.
Emanuel Kanal, MD, FACR Director, MR Services UPMC Presbyterian 200 Lothrop Street Department of Radiology, Room D-132 Pittsburgh, PA 15213-2582MRI (MAGNETIC) COMPATIBILITY OF FLOTEC ALUMINUM REGULATORS
The Flotec Aluminum Regulator is MRI Safe and MRI Compatible.
The following is the test report provided by:
John David, PHD Thermodynamics Director, Engineering Flotec, Inc. 7625 West New York Street Indianapolis, IN 46214
September
19, 1995, updated August 3, 2001
Magnetic attraction is an important factor in MRI (magnetic
resonance imaging) equipment. MRI
equipment have magnets that are so powerful that they can pull steel items
across the room into the magnet. So,
it is important that the magnetic attraction of the items in the room be small.
It is desirable for people on oxygen to bring their breathing
equipment with them into the MRI room. This can be done if the breathing equipment is non-magnetic,
e.g. an aluminum cart, and aluminum tank, a brass or aluminum post valve and an
aluminum regulator. However, our
regulators are not entirely aluminum. Some
critical parts of Flotec regulators are made of “non-magnetic” stainless
steel that becomes magnetic from processing the material.
So, it was necessary to verify how much magnetic attraction our
regulators will experience.
The RR830-100P1 is typical of our aluminum regulators and has
about as much magnetic material as any of our aluminum regulators.
The magnetic attraction of the RR830-100P1 measured at 0.0578 ounces in a
uniform magnetic field of 290 Gauss. Typical
hospital MRI magnets can have a magnetic field as large as 3 Tesla, which is
equivalent to 30,000 Gauss. The
magnetic attraction of the regulator next to a 3 Tesla magnet should calculate
to be about 5.98 oz. assuming a linier relation to static field strength, even
if it were next to the magnet.
In conclusion, the magnetic attraction of Flotec aluminum
regulators is small. There should
be no problem with the regulator being drawn into the magnet of hospital MRI
equipment, especially since the regulator is attached to the post valve and
tank.
|
Test
to determine magnetic attraction in strong uni-directional (DC) magnetic
fields. Test using a friction
free torsion balance to determine attraction in known magnetic field.
Circulated 50 amperes DC in 632 turn, 12 ¼” mean diameter coil
having a measured field strength of 290 gauss at the CG of the Regulator
Positioned above the coil. System
carefully balanced prior to energizing the coil then added weights to
re-balance with attraction of magnetic field.
290 Gauss, Balanced at: 1640 milligrams 1000
gauss =3.4482
x 1.640 Grams = 5.655 Grams at 1000 gauss
290 gauss 1
Gram = 0.03527396 Avdp. Ounces x 5.655 =
0.19947 oz. @ 1000 Gauss Field At
10,000 Gauss Magnetic Attraction = 1.995 oz./10kg. +5%.
Regulator Weights 14 1/8 oz. 1.995
oz.
= 0.41412 or
Attraction = 14% of weight of
regulator. 14.125
oz. |
TB-MRI Compatibility.doc(09-17-01)
This guidance was written prior to the February 27, 1997 implementation of FDA's Good Guidance Practices, GGP's. It does not create or confer rights for or on any person and does not operate to bind FDA or the public. An alternative approach may be used if such approach satisfies the requirements of the applicable statute, regulations, or both. This guidance will be updated in the next revision to include the standard elements of GGP's
A Primer on
Medical Device Interactions with
Magnetic Resonance Imaging Systems
DRAFT
DOCUMENT
This
document is being distributed for comment purposes only.
CDRH
Magnetic Resonance Working Group
Draft
released for comment on: February 7, 1997
The Federal
Register notice reopening the comment period for this document was published May
22, 1997, and extends the comment period to August 20, 1997. Comments and
suggestions regarding this draft document should be submitted to Marlene Skopec,
Office of Science and Technology, HFZ-133, 12721 Twinbrook Pkwy, Rockville, MD
20852. Comments and suggestions received after August 20, 1997, may not be acted
upon by the Agency until the document is next revised or updated. For questions
regarding this draft document, contact Marlene Skopec at (301) 443-3840.
U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Center for Devices and Radiological Health
Table of Contents
|
1.0 |
1 |
|
|
2.0 |
2 |
|
|
3.0 |
4 |
|
|
|
Table 1: MR
Environment Medical Device Concerns |
5 |
|
|
Table 2: Effect of
Medical Device on Operation of MR Scanner |
5 |
|
4.0 |
6 |
|
|
|
4.1 The Static
Magnetic Field and Spatial Gradient |
6 |
|
|
4.2 Pulsed Gradient
Magnetic Fields |
8 |
|
|
4.3 Pulsed Radio
Frequency Fields |
8 |
|
|
4.4 Image Artifacts
and Noise |
10 |
|
5.0 |
11 |
|
|
6.0 |
15 |
|
|
7.0 |
18 |
|
|
8.0 |
19 |
|
|
Appendix A: |
SOME
THOUGHTS ON CRITERIA FOR DEVICES TO BE USED IN THE MR ENVIRONMENT (DRAFT) |
20 |
1.0 Scope/Purpose
With
increasing numbers of medical device manufacturers seeking to make magnetic
resonance (MR) safe or MR compatibility claims for their devices, it is now more
important than ever that medical device reviewers are aware of the potential
implications of these claims. However, use of terms such as "MR
Compatible" and "MR Safe" without reference to the specific MR
environment is vague because the MR environment to which the device was tested
can vary greatly. These claims should be avoided. With the advent of open
magnetic resonance imaging (MRI) systems and interventional MR, the trend of
making MR claims for medical devices will only continue and accelerate.
The purpose of this document is two-fold. It should
serve to sensitize medical device reviewers to the meaning and ramifications of
MR safety or MR compatibility claims. It will also provide a background of MR
theory and the effect the MR environment may have on medical devices. This is
intended to serve as a general background document on medical device
interactions in magnetic resonance imaging systems. It is not intended to
replace documents created that address specific devices or device areas.
Questions or concerns regarding this document may be directed to Marlene Skopec
in OST/DPS/EPB at 443-3840 or by electronic mail at MDS.
Reviewers should direct phone calls
from health care professionals or consumers who wish to make FDA aware of
concerns they have regarding device interactions or other types of problems
associated with the magnetic resonance imaging environment to FDA's voluntary
MedWatch program at 1-800-FDA-1088. If manufacturers, user facility personnel or
device distributors subject to mandatory reporting requirements request a
determination of whether or not an event is required to be reported under the
Medical Device Reporting Regulations, they should be referred to the Office of
Surveillance and Biometrics, Division of Surveillance Systems, MDR
Policy/Guidance Group at (301) 594-2735.
2.0 Definitions
Active
Device:
The
term "active" refers to any medical device that can only serve its
intended use with the supply of power by any means including, but not limited to
line, battery, or gas power. Examples of active devices include ventilators,
pacemakers, and patient monitoring devices.
Five
Gauss Line:
This
line specifies the perimeter around an MR scanner within which the static
magnetic fields are higher than five gauss. Five gauss and below are considered
"safe" levels of static magnetic field exposure for the general
public.
Image
Artifact:
This
is a general term that refers to an inappropriate image signal at a specified
spatial location. It is generally characterized as increased signal intensity in
an area which is known to contain no signal producing material or decreased
signal intensity (voids) where signal should be produced.
*
MR Compatible:
This
term indicates that the device, when used in the MR environment, is MR Safe and
has been demonstrated to neither significantly affect the quality of the
diagnostic information nor have its operations affected by the MR device.
MR
Environment:
This
term is used to describe the general environment present in the vicinity of an
MR Scanner. In particular, this refers to the area within the 5 gauss line
around the scanner. Characteristics of the environment include the following:
1.
the static magnetic field (the range of 0.2 to 1.5 tesla is most common,
but it can exceed 4.0 tesla*) and associated spatial gradients;
2.
rapidly changing magnetic fields (imaging gradients ~kHz); and
3.
radio frequency (RF) magnetic field pulses (on the order of tens to
hundreds of MHz, i.e., in the FM radio band).
* note: 1 tesla = 10,000 gauss
The "MR Environment" includes anywhere in
the MR procedure room, including the center of the bore of the MR scanner.
*
MR Safe:
This
term indicates that the device, when used in the MR environment, hasbeen
demonstrated to present no additional risk to the patient, but may affect the
quality of the diagnostic information.
Passive
Device:
The
term "passive" refers to any medical device that serves its function
without the supply of power. Examples of passive devices include but are not
limited to aneurysm clips, shunts, scalpels, IV poles, and oxygen bottles.
Specific
Absorption Rate (SAR):
SAR
is a measure of the absorption of electromagnetic energy in the body (typically
in watts per kilogram (W/kg)).
Time
Rate of Change of Magnetic Field (dB/dt):
Rate
of change of the magnetic flux density with time (tesla/second).
* The use of the terms, "MR Compatible" and
"MR Safe" without specification of the MR environment to which the
device was tested should be avoided since interpretation of these claims may
vary and are difficult to substantiate rigorously. Statements such as
"intended for use in the MR environment" or similar claims along
with appropriate qualifying information are preferred (i.e. test
conditions should be specifically stated). *
3.0
Introduction and Overview of Medical Device Concerns
Advantages
Unlike x-ray based medical diagnostic techniques such
as computed tomography, magnetic resonance imaging (MRI) and spectroscopy are
techniques which do not employ ionizing radiation. As such, it is considered to
be less hazardous than other x-ray imaging techniques. In addition, since x rays
can only discriminate different tissues by electron density, which does not vary
greatly between soft tissues, the injection of contrast media is often
necessary. In MRI, however, there are a number of tissue specific parameters
which can affect magnetic resonance (MR) signals. One of the most important
advantages of MRI is its capacity for displaying soft tissue contrast. An
example of this capacity is the discrimination between the gray and white matter
of the brain that can be accomplished with MRI. Image contrast can be tailored
to the specific clinical application so that specific types of pathology are
emphasized. In addition, since MRI is unobstructed by bone, it is especially
beneficial in imaging of the brain and spinal cord. MRI also has the unique
ability to acquire images in numerous planes without repositioning the patient.
Three-dimensional recreations of anatomic structure can be obtained. These
characteristics render MRI a very effective and important tool for soft tissue
imaging.
Hazards
The potential benefits of MRI are numerous; however,
there are hazards intrinsic to the MR environment which must be acknowledged and
respected. These hazards may be attributed to one or to a combination of the
three main components that make up the MR environment: a strong static magnetic
field including its associated spatial gradient, pulsed gradient magnetic
fields, and pulsed radio frequency (RF) fields. For a properly operating system,
the hazards associated with direct interactions of these fields and the body are
negligible. It is the interactions of these fields with medical devices placed
within the fields that creates concerns for safety. Each of the component fields
is described in detail along with relevant documented cases of adverse events in
Section 4. There are numerous documented cases of mishaps in the MR environment
that have resulted in injury and even death in a few cases. Those listed are
just a sampling of adverse events that are documented in the Medical Device
Reports (MDR) and Problem Reporting Program (PRP) systems. It is likely that
many adverse incidents occur, but are not reported.
Another aspect of introducing a medical device into
the MR environment is the effect its presence and operation may have on proper
functioning of the MR scanner. Concerns related to this aspect including image
artifact and noise are also addressed greater detail in Section 4.
The tables on this page provide a
summary of the hazards and concerns related to medical devices in the MR
environment.
Table
1: MR Environment Medical Device Concerns
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Table
2: Effect of Medical Device on Operation of MR Scanner
|
|
|
|
|
|
4.0 Components of MRI
and Effects on Medical Devices
The
preceding section provided an introduction and summary of the medical device
concerns related to the MR environment. In this section, each component of the
MR environment is discussed in detail and examples of adverse events are
provided.
4.1
The Static Magnetic Field and Spatial Gradient
An intense static magnetic field is a component of the
MR environment which is always present even when the scanner is not imaging.
This static magnetic field is typically between 0.2 and 2.0 tesla (5,000 to
20,000 Gauss) measured in the center of the magnet bore. Current
state-of-the-art technology is pushing this upper limit to 4 or 5 tesla in
research MRI systems. This is up to 100,000 times the magnetic field strength of
the earth. This strong magnetic field strength drops off rapidly with distance
away from the magnet, producing a large spatial gradient. As a result of this
large gradient, magnetizable objects introduced into the field are accelerated
and can quickly become dangerous projectiles. Imagine a pair of sharp scissors
flying through the air pulled into the magnet bore (see MDRs listed below).
In addition, just as magnetic material aligns itself
with the poles of a permanent bar magnet or a compass needle aligns itself with
the earth's magnetic field, certain objects introduced into the MR environment
will exhibit similar behavior. When brought near the magnet, these objects may
be subjected to a torque which acts to align it with the magnetic field. This
motion can be especially hazardous, for certain implanted medical devices. In
short, the static magnetic field can induce a torque on an object whereas the
associated spatial gradient can exert a translational force on the object. The
magnitude of these effects is dependent on the geometry and mass of the object,
as well as the characteristics of the MR system's magnetic field.
Therefore, patients with certain implanted devices,
such as many types of intracranial aneurysm clips, are contraindicated from MR
imaging since the torque and displacement forces produced on the device can
result in the tearing of soft tissues. In fact, there has been at least one
death recorded as a result of movement of an aneurysm clip. When the event
occurred, the patient was in the process of being moved in towards the magnet,
but had not yet entered the magnet bore. Other implants, such as certain cardiac
pacemakers are known to function erratically even in relatively weak magnetic
fields. In device labeling for pacemakers, MRI is listed as a contraindication.
Individuals with implanted pacemakers, whether or not pacemaker dependent, are
contraindicated from entering the MR procedures room or coming within the 5
gauss line around the scanner. In general, persons with any type of
electrically, magnetically, or mechanically activated implants (pacemakers,
neurostimulators, infusion pumps, etc.), should remain outside the 5 gauss line.
It is important to note that the working of a material
(e.g., machining, molding, bending) may significantly alter its magnetic
properties. This change can be so significant that while the bulk material may
be initially magnetically inert, once it is formed into a medical device, it may
experience torque and translational forces significant enough present a safety
hazard when introduced in the MR environment. Therefore, all testing of devices
for immunity to the strong static magnetic fields should be conducted on the
device in its finished form. Quality assurance is especially important in order
to insure that the behavior of a device in the MR environment does not vary
significantly from item to item.
A
Sampling of MRI Related Incidents from the MDR Database
*
MDR-351516:
A
patient with an implanted cardiac pacemaker died during an MR exam. (12/2/92)
*
MDR-175218:
A
patient with an implanted cardiac pacemaker died during or shortly after an MR
exam. The coroner determined that the death was due to the interruption of the
pacemaker by the MR system. (9/18/89)
MDR-349790:
A
patient with an implanted intracranial aneurysm clip died as a result of an
attempt to scan her. The clip reportedly shifted when exposed to the magnetic
field. The staff apparently had obtained information indicating that the
material in this clip could be scanned safely. (11/11/92)
MDR-100222:
Dislodgement
of an iron filing in a patient's eye during MR imaging resulted in vision loss
in that eye. (1/8/85)
MDR-454660:
A
patient complained of double vision after an MR exam. The MR exam as well as an
x-ray revealed the presence of metal near the patient's eye. The patient was
sedated at the time of the exam and was not able to inform anyone of this
condition. (12/15/93)
MDR-547886:
An
IV pole was attracted to the magnet and struck a patient, cutting his arm. The
patient required stapling of the cut. (8/30/94)
MDR-405200:
A
pair of scissors was pulled out of a nurses hand as she entered the magnet room.
The scissors hit a patient causing a cut on the patient's head. (8/2/93)
MDR-234698:
A
patient was struck by an oxygen bottle while being placed in the magnet bore.
The patient received injuries requiring sutures. (6/2/91)
PRP-19168:
Two
steel tines (parts of a fork lift) weighing 80 pounds each were accelerated by
the magnet striking a technician and knocking him over 15 feet resulting in
serious injury. (6/5/86)
* These events may also be attributed to the pulsed RF
fields.
4.2
Pulsed Gradient Magnetic Fields
Another component of the MR environment is a pulsed
gradient magnetic field that is used for signal localization. When this gradient
magnetic field is applied, the magnetic field intensity changes rapidly, giving
rise to a time-varying magnetic field. During the rise time of the magnetic
field, a voltage is induced in an electrical conductor, even when it is
stationary in the field. However, in most MRI systems, the currents induced by
the pulsed magnetic gradient field are about 1,000 times smaller than those
induced by the pulsed RF component and are therefore not of great concern with
regard to thermal injuries.1 Major concerns with the pulsed gradient
fields are biological effects including electrical nerve stimulation and the
generation of light flashes (magnetophosphenes) that may result from a slight
torque exerted on the retinal cones. Current FDA guidance limits the Time Rate
of Change of Magnetic Field (dB/dt) to levels which do not result in painful
peripheral nerve stimulation.
4.3
Pulsed Radio Frequency Fields
A third main component of the MR
environment is the pulsed radio frequency (RF) magnetic fields which is used to
elicit MR signals from tissue. With regard to biological effects, one main
concern with this component of MR is the production of heat in tissue. The rate
at which RF energy is deposited in tissue is defined as the specific absorption
rate (SAR) which is measured in units of watts per kilogram (W/kg). Current FDA
guidance limits SAR whole body exposure to 4.0 W/kg for patients with normal
thermoregulatory function and 1.5 W/kg for all patients, regardless of their
condition. The duty cycle on the RF pulse during MR imaging is restricted based
on this SAR limit.
With regard to medical devices, electrical currents
may be induced in conductive metal implants, such as skull plates, and hip
prostheses. When conductive patient leads are used during MR scanning, it is
especially critical that no loops are formed by the leads. Looped patient leads
or devices such as the halo device used for spinal immobilization can pick up RF
energy resulting in induced currents, heating of the material, and as a result,
potentially severe patient burns. To further reduce the possibility of burns, it
is recommended to thermally insulate electrically conductive material in the
bore of the magnet from the patient using blankets or sheets.
A
Sampling of MRI Related Incidents from the MDR Database
MDR-711781
An
electrically conductive lead was looped and placed against bare skin causing a
burn on the patient's upper arm. (5/19/95)
MDR-591457:
A
child received a burn to the right hand from an ECG cable while the patient was
anesthetized. A skin graft was required to treat the affected area. (1/26/95)
MDR-246106:
A
patient received a 1.5" x 4" blistered burn to the left side of the
back near the pelvis from an ECG gating cable. (9/23/91)
MDR-701219:
A
patient received blistered burns on the finger where a pulse oximeter was
attached during MR scanning. A skin graft was required to treat the affected
area. (2/27/95)
MDR-391667:
A
patient received small blistered burns to the left thumb and left thigh.
Reportedly, the operator input an inaccurate patient weight resulting in an
incorrect SAR value. (2/10/93)
*
MDR-149476:
A patient with an implanted insulin infusion pump was placed in an MR scanner resulting in movement of the device. The pump was removed from the patient and subsequently found to be non-functional (1/13/88).
* This event may also be attributed
to the static magnetic field.
4.4
Image Artifacts and Noise
Image artifacts and RF noise can be caused by the
presence and/or operation of a medical device in the MR environment. Artifacts
can be caused by medical devices which are in or near the imaging field of view
(such as implants or surface electrodes). Materials produce their own
characteristic static magnetic field that can perturb the relationship between
position and frequency essential to accurate image reconstruction. If the object
has a magnetic susceptibility that is significantly different from that of
tissue, distortion will result. Also, an implant may exhibit an induced eddy
current due to the incident RF magnetic field, altering the RF field near the
implant and thereby causing distortion.
RF noise, which often appears as static on the image,
can be caused by a medical device located anywhere in the MR procedure room. RF
noise is a result of excessive electromagnetic emissions from the medical device
that interfere with the proper operation of the MR scanner. Since the MR
procedure room is shielded from extraneous RF fields entering the room,
operation of electromagnetically noisy equipment outside the room does not
typically affect the MR scanner.
Primary concerns with image artifact and noise include
the production of a void where anatomical information is needed as well as the
production of artifacts that may be misdiagnosed as pathology (see MDR below).
MDR-183091: Surgery was performed on a patient based
on an artifact present on an MR image (1/30/90).
5.0
Basic MR Theory 2,3
The
purpose of this section is to present the interested reader with a general
description of the theory of signal production that is the basis of magnetic
resonance imaging. The components of MR imaging which were introduced in the
above section are described here in greater detail and integrated into the MR
system. The concepts presented here are interesting, but may be skipped by those
not interested in the details of MR theory.
In the nucleus of every atom, individual protons and
neutrons spin about an axis. This property, called spin angular momentum, is the
basis of nuclear magnetism. Since atomic nuclei have charge, this spinning
motion produces a magnetic moment along the spin axis. In most nuclei, the
particles are paired so that the net magnetic properties cancel. However, if the
number of protons or neutrons is odd, complete cancellation is not possible.
Nuclei with an unpaired proton or neutron such as hydrogen 1, carbon 13, and
sodium 23, among others, exhibit a net magnetic effect. The relative strength of
this magnetic moment is a property of the type of nucleus and therefore
determines the MR detection sensitivity. The hydrogen (1H) nucleus,
which is highly abundant in biological systems, has the strongest magnetic
moment.
Since
the individual magnetic moments (or axes of spin) are randomly oriented,
biological tissue does not normally exhibit a net magnetization (Figure 1).2
However, in the presence of an external static magnetic field, B0,
the individual magnetic moments tend to align either parallel or antiparallel to
the direction of the applied field, similar to the way a permanent bar magnet
will align itself with the field or a compass needle aligns with the earth's
magnetic field.
Since
a parallel alignment to the field is the lower energy state, it is preferred and
slightly more nuclei will align parallel rather than antiparallel to the field.
As a result, the tissue will exhibit a net magnetization not unlike like that of
a piece of iron in a magnetic field, although not as strong. The individual
spins do not align exactly parallel to the applied field, but at an angle to it
(Figure 2).2

Like a spinning top, the individual spins cause the
moment to precess about the axis of B0 (Figure 3).2 The
frequency with which the moment precesses is given by the Larmor equation shown
below.
Y B0 = f
where B0 = strength of the applied magnetic field
Y = gyromagnetic ratio (related to
the strength of the magnetic moment for the type of nuclei)
f
= the frequency of precession (Larmor frequency)
For the hydrogen atom Y
= 4257 Hz/Gauss. Therefore, at B0 = 1.5 Tesla (10,000 Gauss = 1
Tesla), the Larmor frequency is 63.855 MHz.
In
order to create an MR signal which can be detected, a resonance condition must
be established. In other words, there must exist a situation of alternating
absorption and dissipation of energy. In the external static magnetic field, B0,
nuclei can be shifted from the parallel to antiparallel alignment by the
application of radio frequency energy. Application of radio frequency (RF)
magnetic field at the Larmor frequency results in energy absorption, while RF
energy applied at other frequencies has no effect. If we consider an RF magnetic
field, B1, applied perpendicular to B0, the system will
absorb energy and begin to precess about the B1 axis (Figure 4). 2
If the RF energy is pulsed, the net magnetization is
rotated to a certain angle away from the B0 axis.This angle is
referred to as the flip angle and is proportional to the duration and amplitude
of the RF pulse. Upon termination of the RF pulse, the nuclei return to their
original alignment parallel to the applied static field and energy is emitted in
the form of a weak RF signal. The frequency of the emitted signal depends on the
strength of the applied static magnetic field as well as the type of nuclei
producing the signal. Detection and analysis of this signal provide insight into
the chemical composition of the material. This process of alternating absorption
and emission of RF energy by the material is termed magnetic resonance
(MR).
At the end of the applied RF pulse, the RF signal
emitted by the material is at its maximum intensity. The signal intensity
diminishes rapidly (within a few hundred milliseconds) as the higher energy
state (the antiparallel state) is depopulated and the nuclei return to their
original energy state. This RF signal is picked up by a receiver coil. The
waveform of this signal is an exponentially damped sine wave and is called the
free induction decay (Figure 5).2

In order to produce an image, each MR signal must be
referenced to a specific region of tissue. This is accomplished by applying a
gradient magnetic field in which the field strength varies linearly with
position. The gradient gradually varies the magnetic field strength resulting in
a corresponding shift in the RF frequency needed to stimulate the tissue. Since
emitted RF signals will also demonstrate a shift in frequency, the excited
tissue from which the signals originated can be localized. Using a
computer-aided reconstruction program, similar to that used in computed
tomography, the signals attributed to individual volume elements of tissue can
be resolved and reconstructed into an image. The most common method of image
reconstruction is the two-dimensional Fourier transform.
6.0 Labeling
*
The use of the terms, "MR Compatible" and "MR Safe" without
specification of the MR environment to which the device was tested should be
avoided since interpretation of these claims may vary and are difficult to
substantiate rigorously. Statements such as "intended for use in the MR
environment" or similar claims along with appropriate
qualifying information are preferred (i.e. test conditions should be
specifically stated). *
Note:
Judgement should be exercised in requesting data for
the substantiation of MR claims. A risk assessment considering the potential
outcomes of adverse interactions in the MR environment should be constructed to
help determine the need for and degree of substantiation necessary for MRI
claims. For example, certain passive devices constructed entirely of polymers
may not require data to substantiate its suitability for use in the MR
environment. The attachment, "Some Thoughts on Criteria for Devices to be
Used in the MR Environment" provides a list of concerns to consider in your
evaluation. For further assistance on specific issues, contact your division MR
Working Group Representative or other group member.
Substantiation of Claims
If a device is to be labeled MR Safe, the
following information should be provided:
a) data demonstrating that when the device is
introduced or used in the MR environment (i.e. the MR scan room) it does not
pose an increased safety risk to the patient or other personnel or;
b) a scientifically based rationale for why data are
not necessary to prove the safety of the device in the MR environment (for
example, a passive device made entirely of a polymer known to be nonreactive in
strong magnetic fields).
If a device is to be labeled MR Compatible, the
following information should be provided:
a) data demonstrating that when the device is
introduced or used in the MR environment, it is MR safe, that it performs its
intended function without performance degradation, and that it does not
adversely affect the function of the MR scanner (e.g. no significant image
artifacts or noise). Any image artifact or noise due to the medical device
should be quantified (e.g., % volume affected, signal to noise ratio); or
b) a scientifically based rationale for why data are
not necessary to prove the compatibility of the device in the MR environment
(see example above).
Claims to include Test Conditions and Outcomes
Claims regarding MR Compatibility, or MR Safety,
including claims stating or implying that a device may be used or brought into
the MR environment, such as "intended for use during MR imaging"
should be substantiated with supporting data. If supporting data is not
provided, the sponsor may provide a justification for why this data is not
necessary to support the claims (e.g., the device is made entirely of a polymer
having a susceptibility similar to tissue). This information should include the
following:
1.
The static magnetic field strength (gauss (G) or tesla (T)) to which the
device was tested and demonstrated to be "safe" ,
"compatible", or "intended for use in" should be related to
typical machine ratings (e.g. 0.5 T, 1.5 T, 2.0 T, and shielded or unshielded
magnet, etc).a
2.
The spatial gradient (field strength per unit distance (i.e., G/cm)) in
which the device was tested and demonstrated to be "safe",
"compatible", or "intended for use in".b
a Unless the device is intended to
be permanently attached to the floor or to the structure of the MR scan room,
the device should be tested at the minimum distance from the magnet isocenter
that it may be physically positioned. If the device can physically fit into the
magnet bore, it should be tested in the bore. The type of magnet used (shielded
or unshielded) should be specified since the magnetic field spatial gradient is
larger in the shielded case.
b The manufacturer should locate the
maximum spatial magnetic field gradient for the MRI system employed and test the
device at that position when practical. The maximum spatial gradients are
typically approximately 240 to 540 G/cm.
3.
The RF transmitter power used during testing of the device.
The above information should be
included on a label conspicuously affixed to the device. This information
including a detailed description of the test conditions applied and a summary of
the test results should be included in associated device literature (e.g.
Operator's Manual). Associated device literature should include the type of
imaging sequence used during testing, a description of artifacts if any caused
by the device, and/or degradation of device function in the MR environment.
Designation of a Separation Distance
Portable devices requiring a separation distance
between the device and the MR magnet should not be considered MR Safe, MR
Compatible, or intended for use in the MR environment. Typically the 5 gauss
line is the only location where the static magnetic field strength is specified
around an MR scanner. Therefore, labeling specifying a separation distance
between the MR magnet and the device to ensure safe or proper operation of the
device should be avoided.
Devices intended to be bolted down or otherwise
permanently affixed to the floor or other unmovable structure may have a
required separation distance. The maximum recommended static field strength and
spatial magnetic gradient field to which the device may be exposed should be
listed on a label that is conspicuously affixed to the device. A separation
distance may be specified only when accompanied with information regarding the
type of magnet (i.e., shielded, unshielded) and static magnetic field strength
and spatial gradient field to which it refers.
Patient Connected Devices, Patient Lead, and
Electrodes
Warnings regarding the potential for heating of
conductive patient connected devices, patient leads, and electrodes that may
result in serious patient burns should be included in the device labeling.
Instructions to the user to help reduce the likelihood of patient burns should
also be included in the device labeling. For example, the user should be warned
not to allow loops in patient leads and not to allow conductive leads come in
contact with bare skin. These warnings should be conspicuously attached to the
device and should also be included in the associated device literature.
Implanted Devices
Since the likelihood of being recommended for an MR
procedure in one's lifetime is increasing, so are the concerns of potential
adverse interactions with implanted devices. Therefore, it is important that
manufacturers of implanted devices identify and address any potential adverse
effects the implant patient may experience as a result of entering high magnetic
field strength areas or undergoing an MR scan. This information should be
provided in the device labeling and associated materials regardless of the
intent to make MR safety or compatibility related claims. The purpose of this is
to inform the clinician and patient of potential concerns with undergoing MR
scanning post implant.
7.0 General Notes on
Testing for MR Compatibility .
Please
note that complete testing of medical devices for MR compatibility requires the
use of a functioning MR imaging system.
Location of Testing
If the subject device is portable (i.e., not
permanently affixed to a specific location in space), the device should be
tested for MR compatibility at the highest static magnetic field strength and to
the largest spatial gradient magnetic field to which it may be physically
exposed. In other words, if the device is small enough to fit in the center of
magnet bore, it should be tested in the bore where the static magnetic field
strength is a maximum. It is important to note that the position of maximum
static field strength for testing of the device will not likely be the same
position of maximum spatial gradient. The position of maximum static and spatial
gradient fields should be located and verified prior to commencement of testing.
For paramagnetic materials and ferromagnetic materials below saturation, the
location where the product of the static magnetic field and the gradient
magnetic field is maximum is important.4
Imaging Sequence
The modes of MR imaging employed during testing should
be representative of the longest expected scanning times and most severe
sequences (e.g., maximum SAR, dB/dt, and RF power). MR compatibility testing of
the device during imaging sequences should be conducted with the device located
at the position of maximum static magnetic field in which it may be physically
placed. In other words, if a portable device is small enough to fit in the
center of the magnet bore, it should be tested in this highest static field
strength.
Effect on the Medical Device
Any effect the MR environment has on the operation of
the device under test should be documented in detail and included in the testing
information provided by the sponsor. Out of specification functioning of the
device while in the MR environment should be explained and demonstrated not to
compromise safety or effectiveness of the device.
Generation of Artifact/Noise
The generation of artifact, RF noise, or other
deleterious effects on the operation of the MR scanner and production of an
image should be documented and included in the testing information provided.
8.0 References
1
"Thermal Injuries and Patient Monitoring during MRI Studies", Health
Devices, Sept. 20(9):362-3, 1991.
2 Keller, P.J., Basic Principles
of Magnetic Resonance Imaging, General Electric Medical Systems, 1991.
3 Jacobson, H.G., Fundamentals of
Magnetic Resonance Imaging, JAMA, 258(23):3417-3423, 1987.
4 Schenck, J.F., "The role of
magnetic susceptibility in magnetic resonance imaging: MRI magnetic
compatibility of the first and second kinds", Med. Phys.
23(6):815-850, 1996.
Appendix
A: SOME THOUGHTS ON CRITERIA FOR
DEVICES TO BE USED IN THE MR ENVIRONMENT (DRAFT)
a) If passive, static magnetic field induced
projectile and attractive problems along with radio frequency (RF) heating
effects may be of concern (see item 2. below).
b) If active, RF and magnetic interference with device
operation can be a concern in addition to item a above.
Consider
that all implants should be labeled regarding their acceptability for use in the
MR environment.
For implants that are known to be hazardous in the MR
environment, consider recommending labeling that suggests that the implanting
physician use a registry service such as Medic Alert.
Revised: M Skopec; 2/4/97